Calibration verification of the International Normalized Ratio

G C Critchfield1, S T Bennett, W R Swaim

  • 1Corning Clinical Laboratories, Teterboro, New Jersey, USA.

American Journal of Clinical Pathology
|December 1, 1996
PubMed
Summary

Laboratories can now locally verify the accuracy of their International Normalized Ratio (INR) results using a new method with standard plasmas. This approach helps ensure reliable prothrombin time reporting across different testing systems.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The immature platelet fraction: creating neonatal reference intervals and using these to categorize neonatal thrombocytopenias.

Journal of perinatology : official journal of the California Perinatal Association·2017
Same author

The iron status at birth of neonates with risk factors for developing iron deficiency: a pilot study.

Journal of perinatology : official journal of the California Perinatal Association·2016
Same author

Comparing automated vs manual leukocyte differential counts for quantifying the 'left shift' in the blood of neonates.

Journal of perinatology : official journal of the California Perinatal Association·2016
Same author

Both initial red cell distribution width (RDW) and change in RDW during heart failure hospitalization are associated with length of hospital stay and 30-day outcomes.

International journal of laboratory hematology·2016
Same author

Reference intervals for reticulocyte parameters of infants during their first 90 days after birth.

Journal of perinatology : official journal of the California Perinatal Association·2015
Same author

Whole-blood viscosity in the neonate: effects of gestational age, hematocrit, mean corpuscular volume and umbilical cord milking.

Journal of perinatology : official journal of the California Perinatal Association·2013

Area of Science:

  • Clinical Chemistry
  • Hematology
  • Laboratory Medicine

Background:

  • The International Normalized Ratio (INR) system aims to standardize prothrombin time (PT) reporting across diverse reagent and instrument platforms.
  • Complete standardization is hindered by inaccurate International Sensitivity Index (ISI) assignments and local reagent/instrument variations.
  • Currently, laboratories lack an easy method to independently verify the accuracy of their INR measurements.

Purpose of the Study:

  • To present a novel method enabling individual laboratories to locally verify the accuracy of their INR results.
  • To demonstrate a practical approach for assessing the reportable range of INR accuracy within a laboratory setting.

Main Methods:

  • Utilized standard lyophilized plasmas with INR values calibrated against the International Reference Preparation (IRP) RBT/90 rabbit thromboplastin.

Related Experiment Videos

  • Illustrated the method on a single coagulometer using two thromboplastin lots with different ISI values (1.20 and 1.97).
  • Main Results:

    • The developed method allows for the local establishment of a reportable accuracy range for INR measurements at a defined error tolerance.
    • Demonstrated superior accuracy with Thromboplastin-DS (ISI 1.20) compared to Thromboplastin-D (ISI 1.97) in the illustrated example.
    • The method is applicable to laboratories of all sizes for calibration verification of INR reagent/instrument systems.

    Conclusions:

    • This method empowers laboratories to perform essential calibration verification for their INR testing systems.
    • Implementing this alongside precision assessments can aid in selecting superior reagent/instrument systems.
    • Ultimately, this leads to more accurate and clinically relevant INR results, improving patient care.