Related Experiment Videos
Migration of press-fit cups fixed with poly-L-lactic acid or titanium screws: a randomized study using
J Thanner1, J Kärrholm, H Malchau
1Department of Orthopaedics, Sahlgrenska University Hospital, Göteborg, Sweden.
Abstract:
The early migration of porous acetabular cups was studied in 43 patients with osteoarthrosis. The patients were randomly allocated to additional fixation of the cup with either biodegradable poly-L-lactic acid screws or titanium screws. Radiostereometric evaluation was done during the first 2 years after the procedure in 43 hips (23 with poly-L-lactic acid screws and 20 with titanium screws). At the 2-year follow-up, cups fixed with poly-L-lactic acid screws had migrated significantly more in the proximal-distal (p < 0.05) and medial-lateral (p < 0.05) directions. Cups with titanium screws displayed more pronounced rotations around the longitudinal axis (p < 0.05). Postoperatively, on the lateral view, there was an increased occurrence of radiolucencies at the dome of the cups fixed with poly-L-lactic acid screws (p < 0.05). The clinical result did not differ between the two groups. Inferior implant-bone contact in the poly-L-lactic acid group, local changes of the bone quality, and diminishing support of the poly-L-lactic acid screws caused by their degradation with time could be reasons for the different pattern of migration observed.