Related Experiment Videos
Configuration of the SCOTI device with different tracheal tubes
N Nandwani1, R Caranza, E S Lin
1University Dept of Anaesthesia, Leicester Royal Infirmary.
Anaesthesia
|October 1, 1996
Abstract:
We have evaluated the Sonomatic Confirmation of Tracheal Intubation device (SCOTI) by testing its ability to be correctly configured with a variety of tracheal tubes of differing internal diameter and length. The device only configured correctly for RAE tubes with internal diameter of 7.0 mm or greater and for armoured tubes of internal diameter 8.5 mm. For conventional tubes of varying internal diameter cut to different lengths, configuration was only successful with certain dimensions. The inability to configure the device correctly with all types and lengths of tracheal tubes limits its usefulness as a indicator of tracheal intubation.