A second defibrillator chest patch electrode will increase implantation rates for nonthoracotomy defibrillators
A J Solomon1, J F Swartz, D J Rodak
1Department of Medicine, Georgetown University Medical Center, Washington, D.C., USA.
Pacing and Clinical Electrophysiology : PACE
|September 1, 1996
Summary
Adding a second chest patch electrode improves nonthoracotomy defibrillator implantation success. This strategy lowers defibrillation thresholds in patients with high thresholds, ensuring successful system implantation.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Nonthoracotomy defibrillator systems offer lower morbidity and mortality than epicardial systems.
- Implantation failure rates for nonthoracotomy systems can reach up to 15%, particularly with monophasic waveforms.
- Elevated defibrillation thresholds pose a significant challenge to successful nonthoracotomy defibrillator implantation.
Purpose of the Study:
- To prospectively evaluate the effectiveness of incorporating a second chest patch electrode in nonthoracotomy defibrillator systems.
- To determine if an additional electrode can overcome high defibrillation thresholds and improve implantation success rates.
Main Methods:
- Fourteen patients with elevated defibrillation thresholds (≥ 24 J) were studied.
- An initial system with a right ventricular electrode and a single subscapular chest patch electrode was used.
- A second chest patch electrode was added (anterior subcutaneous or subscapular submuscular) if the initial threshold was high.
Main Results:
- The addition of a second chest patch electrode reduced system impedance from 72.3 ± 13.3 Ω to 52.2 ± 8.6 Ω.
- Defibrillation thresholds decreased significantly from ≥ 24 J to 16.6 ± 2.8 J with the two-patch system.
- All fourteen patients achieved successful implantation of the nonthoracotomy defibrillator system.
Conclusions:
- A second chest patch electrode, particularly with a subscapular placement, effectively lowers defibrillation thresholds.
- This strategy significantly increases implantation success rates for nonthoracotomy defibrillator systems in challenging cases.
- The findings support the use of an additional chest patch electrode to optimize nonthoracotomy defibrillator system performance.