Related Experiment Video
Updated: Sep 17, 2026

A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
Hydraulic assessment of the floating impeller phenomena in a centrifugal pump
T Nakazawa1, Y Takami, K Makinouchi
1Baylor College of Medicine, Department of Surgery, Houston, Texas 77030, USA.
Abstract:
A compact eccentric inlet port centrifugal blood pump (C1E3) has been perfected for a long-term centrifugal ventricular assist device as well as a cardiopulmonary bypass pump. The C1E3 pump incorporates a sealless design and a blood stagnation free structure. The pump's impeller is magnetically coupled to the driver magnet in a sealless manner. The latest hemolysis study reveals that hemolysis is affected by the magnetic coupling distance between the driver and impeller magnet. Furthermore, a floating phenomenon can be observed in a pivot bearing supported pump. Attention was focused on the relationship between the floating phenomenon's characteristics and the magnetic coupling design in the C1E3 pump. Studies were conducted to evaluate the hydromechanical performance in the floating phenomenon. In this study, the relationship between the magnetic coupling design and the floating phenomenon was verified with a smooth spinning condition. The optimized magnetic coupling distance for the floating mode was estimated to be 12 mm for left ventricular assist device and 9 mm for cardiopulmonary bypass pump. Obtaining an optimal spinning condition is required for regulating the magnetic coupling force. To develop a double pivot bearing pump, it is necessary to establish an optimal spinning and/or floating condition and to determine the proper magnetic coupling and magnetic force between the impeller and driver.
Related Concept Videos
Application of Pascal's Law
Hydraulic systems are used to operate automotive brakes, hydraulic jacks, and numerous other mechanical systems. We can derive a relationship between the forces in a simple hydraulic system by applying...
Fluid Pressure
According to Pascal's law, a fluid at rest will generate equal pressure in all directions. This pressure is measured as a force per unit area, and its magnitude depends on the fluid's specific weight or...
Application of the Energy Equation
Design Example: Creating a Hydraulic Model of a Dam Spillway
Hydraulic Jump
Hydraulic Jump: Problem Solving