Identifying a new reality: zero vascular access recirculation using ultrasound dilution
J T MacDonald1, M A Sosa, N M Krivitski
1Transonic Systems, Inc. Ithaca, NY, USA.
Summary
Blood urea nitrogen (BUN) methods for measuring dialysis access recirculation are unreliable. Ultrasound dilution, a new method, reveals most patients have zero access recirculation, especially with adequate blood flow.
Area of Science:
- Nephrology
- Vascular Access
- Dialysis Technology
Background:
- Blood urea nitrogen (BUN) sampling is a common method for assessing dialysis access recirculation.
- Concerns exist regarding the accuracy and reliability of BUN measurements, particularly at low recirculation levels.
- Recent studies suggest lower recirculation rates than previously assumed in hemodialysis patients.
Purpose of the Study:
- To evaluate the reliability of a novel ultrasound dilution method for measuring access recirculation.
- To investigate the relationship between access flow and recirculation.
- To challenge the existing understanding of access recirculation in hemodialysis.
Main Methods:
- Utilized ultrasound dilution technique to measure access recirculation.
- Compared ultrasound dilution results with traditional BUN sampling methods.
- Analyzed data to determine the correlation between vascular access flow rates and recirculation.
Main Results:
- Ultrasound dilution method demonstrated superior accuracy, avoiding limitations of BUN measurements.
- The study identified a significant proportion of patients with zero access recirculation.
- A strong correlation was found between adequate vascular access flow and the absence of recirculation.
Conclusions:
- Ultrasound dilution is a reliable method for assessing access recirculation.
- The clinical reality of zero access recirculation is supported by new data.
- Maintaining adequate vascular access flow is crucial for preventing recirculation in hemodialysis patients.