Comparative pull-out strength of tapped and untapped pilot holes for bicortical anterior cervical screws
J F Ronderos1, R Jacobowitz, V K Sonntag
1Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St. Joseph's Hospital and Medical Center, Phoenix, Arizona, USA.
Study Design:
This biomechanical study analyzed the axial pull-out strength of tapped versus untapped pilot holes for bicortical screws in the anterior cervical spine.
Objective:
To determine which pilot hole preparation method was mechanically better.
Summary Of Background Data:
Tapping pilot holes in the lumbar spine was previously shown significantly to reduce pull-out strength of pedicle screws. No study was found investigating the effect of tapping on pilot holes for anterior cervical bicortical screws.
Methods:
Twenty-five unembalmed human cadaveric cervical vertebrae (C3-C7) were tested. Two identical pilot holes were drilled into each vertebra: one pilot hole was tapped, and the control pilot hole was not tapped. A fully threaded cortical bone screw was inserted into each pilot hole. Screw pull-out strength was determined using a servocontrolled hydraulic materials testing system and an axial load cell. Force-deformation and failure curves were obtained.
Results:
There were no statistically significant differences between the axial pull-out strength of tapped and untapped pilot holes at any vertebral level. Mean force to-failure was 386 +/- 42 N in the untapped pilot holes and 397 +/- 48 N in the tapped pilot holes.
Conclusions:
Tapping a pilot hole for bicortical screws of the anterior cervical spine neither weakens nor strengthens the axial pull-out strength of fully threaded cortical bone screws. Tapping may be unnecessary; however, it may be desirable in patients with dense bone to cut the thread profile into the bone or if the screws have dull tips and threads.