Inappropriate rate response in a VVI-R pacemaker

M Vanderheyden1, W Timmermans, M Goethals

  • 1Cardiovascular Center, O.L.V. Ziekenhuis, Aalst, Belgium.

Acta Cardiologica
|January 1, 1996
PubMed
Summary

Rate-adaptive pacemakers can sometimes accelerate heart rate inappropriately during upper limb movements. This issue, related to minute ventilation sensors, can be resolved by repositioning the pacemaker can.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Aortic vascular graft infection due to Cardiobacterium Hominis in a heart transplant recipient.

Acta cardiologica·2020
Same author

Heart failure with preserved ejection fraction or non-cardiac dyspnea in paroxysmal atrial fibrillation: The role of left atrial strain.

International journal of cardiology·2020
Same author

In vitro performance of combinations of anti-siphon devices with differential pressure valves in relation to the spatial position.

Acta neurochirurgica·2020
Same author

Wasted septal work in left ventricular dyssynchrony: a novel principle to predict response to cardiac resynchronization therapy.

European heart journal. Cardiovascular Imaging·2016
Same author

Electrical remodeling reflected by QRS and T vector changes following cardiac resynchronization therapy is related to survival in heart failure patients with left bundle branch block.

Journal of electrocardiology·2015
Same author

Epicardial catheter-based ventricular reconstruction (ECVR) in a patient with ischemic heart failure and an anteroapical aneurysm.

Acta clinica Belgica·2014

Area of Science:

  • Cardiology and Biomedical Engineering
  • Pacemaker Technology and Physiological Mimicry

Background:

  • Ongoing efforts seek pacemakers that accurately mimic normal physiological function.
  • Rate-adaptive pacing offers hemodynamic advantages over classical ventricular pacing.
  • The ideal sensor for physiological demands in pacemakers remains elusive.

Observation:

  • A case study involving a VVI-R pacemaker utilizing minute ventilation sensing.
  • The patient exhibited an unexpected heart rate increase during upper limb movements.
  • This acceleration was attributed to transthoracic impedance changes, not sensor malfunction.

Findings:

  • Inappropriate heart rate acceleration in VVI-R pacemakers with minute ventilation sensors is a known complication.
  • Changes in transthoracic impedance during upper limb exercise can trigger this response.
  • The specific patient's issue was resolved by medially repositioning the pacemaker can.

Implications:

  • Awareness of sensor-related issues in rate-adaptive pacemakers is crucial for clinicians.
  • Pacemaker placement may influence sensor performance and patient outcomes.
  • Further research into robust sensor technology and implantation techniques is warranted.