A real vessel phantom for imaging experimentation

W Dabrowski1, J Dunmore-Buyze, R N Rankin

  • 1Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada.

Medical Physics
|May 1, 1997
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Reliability and concurrent validity of three-dimensional ultrasound for quantifying knee cartilage volume.

Osteoarthritis and cartilage open·2022
Same author

High tibial osteotomy to neutral alignment improves medial knee articular cartilage composition.

Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA·2021
Same author

Extending the dynamic range of biomedical micro-computed tomography for application to geomaterials.

Journal of X-ray science and technology·2019
Same author

Poster - Thurs Eve-02: 3 dimensional ultrasound-guided breast brachytherapy.

Medical physics·2017
Same author

Poster - Thurs Eve-12: A needle-positioning robot co-registered with volumetric x-ray micro-computed tomography images for minimally-invasive small-animal interventions.

Medical physics·2017
Same author

Sci-Fri AM: YIS-05: A new guidance device for lateral-approach stereotactic breast biopsy.

Medical physics·2017
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

This study introduces a novel vascular phantom using real human vessels for improved imaging evaluation. This realistic model enhances the accuracy of ultrasound and CT imaging for various vascular conditions.

Area of Science:

  • Biomedical Engineering
  • Medical Imaging
  • Vascular Research

Background:

  • Vascular phantoms are crucial for evaluating imaging modalities like ultrasound (US) and CT.
  • Existing phantoms often lack the precise geometry and pathological features of human vasculature.
  • Previous models, including latex tubes and agar molds, had limitations in anatomical accuracy.

Purpose of the Study:

  • To develop and evaluate a novel vascular phantom utilizing a real human vessel.
  • To assess the phantom's efficacy in mimicking human vasculature, including pathologies like plaques and calcifications.
  • To compare imaging results from the novel phantom using multiple modalities.

Main Methods:

  • A phantom was constructed using a formaldehyde-fixed human vessel section within an agar-filled acrylic box.

Related Experiment Videos

  • The vessel was cannulated onto acrylic tubes to simulate blood flow pathways.
  • Imaging evaluation was performed using X-ray angiography, CT, and 3D B-mode US.
  • Main Results:

    • The phantom demonstrated good correlation with anatomical features such as lumen, plaques, and calcifications across imaging modalities.
    • X-ray angiography, CT, and ultrasound images showed consistent representation of the vessel's geometry and pathology.
    • Minor discrepancies and artifacts were identified and discussed, providing insights for future improvements.

    Conclusions:

    • The real vessel phantom offers a more anatomically accurate and pathologically representative model compared to previous phantoms.
    • This advanced phantom is highly suitable for evaluating and validating medical imaging techniques, including projection radiography, CT, and US.
    • The potential exists to extend its application to magnetic resonance (MR) and US-based flow studies.