Indirect calorimetry during treadmill walking--a study of two methods
M Ohrström1, C Holmer, M Larsson
1Department of Physical Therapy, University Hospital, Lund, Sweden.
Clinical Physiology (Oxford, England)
|May 1, 1997
Summary
Measuring energy expenditure during treadmill walking is crucial for physiotherapy. The Sensormedics 2900 device accurately assessed energy expenditure and respiratory quotient, unlike the Deltatrac monitor, making it suitable for clinical use.
Area of Science:
- Physiology
- Biomechanical Engineering
- Rehabilitation Science
Background:
- Physiotherapy evaluation of walking requires objective measures.
- Energy expenditure measurement offers a valuable supplement to assess treatment outcomes.
- Treadmill walking provides a controlled environment for assessing energy expenditure.
Purpose of the Study:
- To assess the reproducibility of oxygen consumption measurements using two devices during treadmill walking.
- To determine if measurements are sensitive enough to detect differences in energy expenditure and respiratory quotient (RQ) at varying walking speeds.
Main Methods:
- Test-retest reliability study involving two metabolic monitors: Deltatrac and Sensormedics 2900.
- Subjects walked on a treadmill at comfortable and preset speeds for 10 minutes to achieve steady-state measurements.
- Measurements of oxygen consumption and RQ were repeated within two weeks.
Main Results:
- Both devices demonstrated acceptable repeatability for measuring energy expenditure and RQ.
- Both methods were sensitive to differences in energy expenditure across various walking speeds.
- The Deltatrac monitor yielded excessively low RQ values, likely due to airflow issues.
Conclusions:
- The Sensormedics 2900 is a reliable and sensitive tool for evaluating energy expenditure during treadmill walking.
- The Deltatrac monitor's limitations in RQ measurement make it unsuitable for this application.
- Objective energy expenditure data can enhance the evaluation of physiotherapy interventions for walking.