Related Experiment Videos
Rapid liquid-liquid extraction of cocaine from urine for gas chromatographic-mass spectrometric analysis
D Garside1, B A Goldberger, K L Preston
1Diagnostic Referral Laboratories, University of Florida College of Medicine, Gainesville 32608, USA.
Abstract:
A novel, simple and economic liquid-liquid extraction method for isolating cocaine from urine was developed utilizing gas chromatography-mass spectrometry (GC-MS) for analysis and quantification. The use of a single nonpolar organic solvent allowed only nonpolar analytes to be extracted from the biological fluid, and consequently, no derivatization step was necessary before GC-MS analysis. Large numbers of specimens (>60) can be extracted in approximately 3 h with this procedure. The method is highly precise (C.V. <7%), accurate (>98%), sensitive (limit of detection of 5 ng/ml) and has a mean recovery of 48.8%.