[Highly reproducible technique for the estimation of trabecular bone density: three slice method on peripheral
K Mizukawa1, N Sugiyama, Y Mori
1Department of Radiology, Mizushima Kyodo Hospital.
Abstract:
The reproducibility of bone mineral density (BMD) measurements with peripheral quantitative CT has been limited by the repositioning error. In this study, 1 mm-step 3-slice scan data were used to compensate for this error in trabecular BMD of the distal radius. The assessment was based on the liner relations between the cross-sectional area and trabecular bone mineral content and trabecular area ratio, on the condition that the trabecular bone is defined by a BMD value. The estimated reproducible errors of less than 2% under clinical conditions indicate that the method is reliable for follow-up examination.