Fracture embolization of a Duromedics mitral prosthesis
Related Experiment Video
Updated: Nov 1, 2026

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
F J Baumgartner1, A I Munro, W R Jamieson
1Division of Cardiovascular and Thoracic Surgery, Vancouver General Hospital, University of British Columbia, Canada.
Abstract:
The Duromedics bileaflet pyrolitic carbon mechanical prosthesis was introduced by Hemex in 1982 and subsequently acquired by Baxter. This communication documents a case of sudden leaflet fracture of a Duromedics mitral valve 48 months after implantation, which was managed successfully by replacement with a St. Jude Medical mechanical prosthesis. The patient presented in acute distress with paroxysmal atrial tachycardia and pulmonary edema. Transesophageal echocardiography was used to diagnose the leaflet fracture. The fracture had occurred transversely, with the fragments embolizing bilaterally to the iliofemoral arteries. These were removed at a subsequent operation. Cases of such fractures of the Duromedics prosthesis have been reported, with cavitation damage being the postulated mechanism.
Related Concept Videos
Mitral Stenosis III: Medical Management
Mitral Regurgitation III: Medical Management