The MDA-180 coagulation analyser: a laboratory evaluation

A Mohammed1, P A Mehrabani, R Coombs

  • 1Department of Hematology, Westmead Hospital, NSW Australia.

Pathology
|May 1, 1997
PubMed
Summary

The MDA-180 automated coagulation analyser demonstrates reliability and ease of use in a diagnostic laboratory setting. It offers fast sample throughput and accurate analytical performance for various hemostasis assays.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Nutrition in Infants with Orofacial Clefts: An Assessment of Feeding Practices and Factors Affecting Feeding.

Nigerian journal of clinical practice·2026
Same author

THE HEPATITIS B AND HEPATITIS C VIRUSES SEROPREVALENCE AMONG BLOOD DONORS IN AL-DAMAZIN STATE, SUDAN: A THREE-YEAR RETROSPECTIVE STUDY.

Georgian medical news·2025
Same author

Echocardiography Training Pathway for Intensive Care Trainees: A Comparison with International Guidelines.

Irish medical journal·2025
Same author

CARCINOMA-EX- PLEOMORPHIC ADENOMA ARISING WITHIN A PALATAL MINOR SALIVARY GLAND: A CASE REPORT AND REVIEW OF LITERATURE.

West African journal of medicine·2024
Same author

A Comparative Study of Intelligence Quotients of Almajirai Attending Quranic Schools and Pupils of Selected Primary Schools in Zaria, Northwest Nigeria.

West African journal of medicine·2024
Same author

PARTIAL PURIFICATION OF GLUTATHIONE PEROXIDASE ENZYME FROM WOMEN WITH BREAST CANCER.

Georgian medical news·2024

Area of Science:

  • Clinical diagnostics
  • Hematology
  • Laboratory automation

Background:

  • The MDA-180 automated coagulation analyser was evaluated in a routine diagnostic hemostasis laboratory.
  • The instrument has been in use for over 18 months, processing more than 22,000 samples.

Purpose of the Study:

  • To assess the technical performance, ease of use, reliability, and analytical capabilities of the MDA-180.
  • To evaluate its ability to perform assays currently done with alternative methods and its potential for streamlining workflows.
  • To assess its compatibility with laboratory information systems.

Main Methods:

  • The MDA-180 was used in a working diagnostic laboratory environment.
  • Over 22,000 samples were processed, with >90% using the cap-piercing function.

Related Experiment Videos

  • Co-evaluation of assays including PT/INR, APTT, TT, Fibrinogen, Protein C, ATIII, coagulation factors, Lupus anticoagulant (dRVVT), and heparin (alpha Xa).
  • Testing of various reagents, particularly for PT and APTT assays.
  • Interfacing with the CERNER PATHNET Laboratory Information System.
  • Main Results:

    • The MDA-180 demonstrated remarkable intra-assay and inter-assay precision for PT and APTT.
    • Comparative assessments showed good correlation with existing test assay systems.
    • The instrument was successfully interfaced with the CERNER PATHNET system.
    • Over 90% of samples were processed using the automated cap-piercing facility.

    Conclusions:

    • The MDA-180 automated coagulation analyser is reliable and user-friendly.
    • It is capable of fast sample throughput and accurate analysis of hemostasis assays.
    • The instrument is suitable for use in a diagnostic hemostasis laboratory environment.