Related Experiment Video
Updated: Jul 6, 2026

Assessment of Audio-Tactile Sensory Substitution Training in Participants with Profound Deafness Using the Event-Related Potential Technique
Published on: September 7, 2022
[Hearing impairment fitting: study of an auditory prosthesis with 7 entirely programmable filters]
1Service ORL, Hôpital Saint-Antoine, Paris.
Abstract:
We designed a 7-filter digital auditory prosthesis table prototype. For each of these filters frequency band width, amplification and compression were programmable in order to adapt these parameters to the deaf patient's audiometric particularities. We compared the hearing improvement it was possible to obtain with our prototype as a function of the number of filters (3, 4 or 7) and their frequency band width programmability. We tested 21 patients suffering from middle or severe neurosensory hearing loss. This study allows to demonstrate that a 7 programmable-width-filter strategy seems to be the most appropriate.