Flow-proportional administration of nitric oxide with a new delivery system: inspiratory nitric oxide concentration fluctuation during different flow conditions
Related Experiment Video
Updated: Oct 9, 2026

Parallel-plate Flow Chamber and Continuous Flow Circuit to Evaluate Endothelial Progenitor Cells under Laminar Flow Shear Stress
Published on: January 17, 2012
M Sydow1, F Bristow, J Zinserling
1Department of Anesthesiology, UT Houston Medical School, USA.
Objective:
To evaluate the accuracy of a flow-proportional delivery system and the pattern of inspiratory nitric oxide (NO) concentration during different flow conditions.
Design:
Laboratory study in a lung model.
Setting:
University experimental laboratory.
Subject:
With a new delivery system, NO was administered proportional to the inspiratory flow into the inspiratory circuit to deliver a NO concentration of 10 and 30 ppm to a test lung during different ventilatory modes (volume-controlled ventilation [VCV], pressure-controlled ventilation [PCV], and airway pressure release ventilation [APRV]) with a fraction of inspired oxygen (FIO2) of 1.0.
Interventions:
During VCV and PCV, the flow pattern was varied to achieve tidal volumes of 300, 600, and 900 mL, respectively, with inspiratory to expiratory time ratios of 1:3, 1:2, and 1:1. APRV was studied at a minute volume of 6, 12, and 18 L. Nitric oxides (NOx [NO+NO2]) and nitric dioxide (NO2) were monitored by chemiluminescence and electrochemical analysis. As the NO/N2 gas mixture is the only nitrogen source during ventilation with an FIO2 analyzer.
Results:
During all flow conditions, NOx concentration was stable but slightly higher than expected. Measured and expected mean concentrations differed <9% (mean, <4%). Inspiratory NOx concentration fluctuation derived from N2 concentration was significantly higher than expected at higher flow rates, but this difference was not detected by chemiluminescence or electrochemical analysis. The NO2 production was not affected by the flow rate and was always < or =0.2 ppm (NO, 10 ppm) and < or =1.9 ppm (NO, 30 ppm).
Conclusion:
The tested NO delivery module administered stable mean inspiratory NO concentrations. Although inspiratory NO concentration fluctuates depending on the inspiratory flow rate, this delivery device allows stable NO administration without requiring adjustments when ventilator settings are changed.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
10:12Measurement of Nitric Oxide Production in Mouse Tracheal Epithelial Cell Cultures Differentiated at the Air Liquid Interface
Published on: July 3, 2026
Related Concept Videos
Nitric Oxide Signaling Pathway
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Pipe Flowrate Measurement
The orifice meter is a simple,...