Early failure of the femoral component in unicompartmental knee arthroplasty
G D Riebel1, F W Werner, D C Ayers
1Schenectady Orthopaedic Associates, PC, Schenectady, New York 12308, USA.
Abstract:
Unicompartmental knee arthroplasty has been proposed as treatment for unicompartmental arthritis. Overall, results have been satisfactory, with occasional failures of tibial component fixation. Short-term loosening of the femoral components in unicompartmental knee arthroplasty that required revision to total knee arthroplasty is reported. All patients presented with pain and instability of the knee, with loosening confirmed radiographically and at revision surgery. The Porous-Coated Anatomic femoral component (Howmedica, Rutherford, NJ) was biomechanically tested after being implanted in cadaver limbs. Cyclic loading in flexion and extension under physiologic loads resulted in implant failure. The mode of failure was rocking in the sagittal plane, which corresponds to the clinical findings. The femoral implant design consists of an angled bone-implant interface with straight arms, which causes shear at the bone-prosthesis interface did not loosen under identical biomechanical testing in paired knees. It was concluded that the Porous-Coated Anatomic femoral component has an unacceptably high early clinical loosening rate, confirmed by comparative biomechanical testing.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023