Biliary stricture measurement and stent selection
1Department of Medicine, Robert C. Byrd Health Science Center, West Virginia University, Morgantown 26506-9161, USA.
Objective:
Because of variations in magnification, errors in stent selection may occur when stricture location is determined from fluoroscopic images or x-ray film.
Methods:
An ERCP catheter with measurement markings was inserted 5, 7, and 10 cm into the bile duct in 30 patients without bile obstruction. Film measurements obtained at each depth were converted to actual distances using standard conversion as well as endoscope ratio conversion formulas. The site of obstruction in 52 patients with an obstructing lesion was measured with a ruled catheter, by a wire withdrawal technique, and using x-ray film. After a stent was selected on the basis of the catheter measurement, the accuracy of stent selection was determined for each method of measure.
Results:
Conversion of x-ray measurements to ruled catheter measurements obtained by insertions of 5, 7, and 10 cm using standard conversion factors yielded measurements of 4.9 +/- 0.9, 6.8 +/- 0.3, and 9.5 +/- 1.9 cm, respectively, with a correlation coefficient of r = 0.80. Ratio conversion yielded measurements of 0.5 +/- 0.8, 7 +/- 1.0, and 9.9 +/- 1.4, respectively, with a correlation coefficient of r = 0.88. Measurement of stricture location with the ruled catheter and then by wire withdrawal yielded a correlation coefficient of 0.98. When ruled catheter measurements were compared with the x-ray ratio conversions, the concordance dropped to 0.79. The ruled catheter and wire withdrawal were more accurate in predicting the location of the stent tip than x-ray film measurements (p < 0.001, Wilcoxon matched pairs). Of 52 stents selected, no errors in stent selection occurred when the ruled catheter was used (p < 0.001, Fisher's exact test), two errors occurred when wire withdrawal was used (p < 0.004), and 14 errors occurred when film measurements were used.
Conclusions:
Use of a ruled catheter or wire withdrawal is much more accurate for selecting stents than use of x-ray film measurements.
Related Concept Videos
Esophageal Strictures-I: Introduction
Etiology
The primary cause of esophageal strictures is long-standing gastroesophageal reflux disease (GERD), accounting for about 70 to 80% of adult cases. Chronic acid reflux can lead to injury and scarring of the esophageal lining, culminating in...
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Inflammatory Bowel Disease V: Surgical Management
Here are some common surgical interventions for IBD: