Summary
Directly coupling micellar electrokinetic chromatography (MEKC) with mass spectrometry (MS) offers enhanced analysis. Various on-line coupling strategies mitigate surfactant interference, enabling robust identification and structure confirmation.
Area of Science:
- Analytical Chemistry
- Separation Science
- Mass Spectrometry
Background:
- Micellar electrokinetic chromatography (MEKC) coupled with mass spectrometry (MS) is valuable for analyte identification and structure confirmation.
- Direct coupling faces challenges due to nonvolatile MEKC surfactants affecting MS performance, causing sensitivity loss and ion source contamination.
Purpose of the Study:
- To review and evaluate various on-line coupling approaches for MEKC-MS.
- To identify methods that overcome the limitations of direct MEKC-MS coupling.
Main Methods:
- Review of different on-line MEKC-MS coupling techniques.
- Evaluation of strategies including high-molecular-mass surfactants, electrospray-chemical ionization (ES-CI) interfaces, voltage switching, buffer renewal, partial-filling micellar plugs, and anodically migrating micelles.
Main Results:
- The ES-CI interface shows promise for routine on-line MEKC-MS with nonvolatile salts and surfactants.
- High-molecular-mass surfactants reduce background ions and allow low surfactant concentrations.
- Partial-filling micellar plugs and anodically migrating micelles effectively prevent MEKC micelles from entering the ESI-MS system.
Conclusions:
- On-line MEKC-MS coupling can be achieved without instrument modification using partial-filling micellar plugs or anodically migrating micelles.
- The ES-CI interface is a promising solution for routine MEKC-MS operations.
- These advancements facilitate direct transfer of MEKC separations to ESI-MS.