Inappropriate shocks in patients receiving internal cardioverter-defibrillators for malignant ventricular arrhythmias

K Prasad1, A G Kishore, M H Anderson

  • 1Department of Cardiological Sciences, St. George's Hospital Medical School, London.

Insights

Internal cardioverter defibrillators (ICDs) can deliver inappropriate shocks, often due to atrial fibrillation. This study highlights the frequency of these shocks in patients receiving ICDs for malignant ventricular arrhythmias.

Area of Science:

  • Cardiology
  • Medical Devices

Background:

  • Implantation of internal cardioverter defibrillators (ICDs) is a treatment for malignant ventricular arrhythmias.
  • Therapy failure and inappropriate shocks are known complications of ICDs.

Purpose of the Study:

  • To investigate the incidence and causes of shocks in patients receiving ICDs.
  • To assess the long-term outcomes of ICD implantation.

Main Methods:

  • Study included 81 patients undergoing ICD implantation.
  • Data collected via Holter monitoring, telemetry, device memory, and clinical follow-up.
  • Follow-up duration was 36 months for 58 patients.

Main Results:

  • 35 patients experienced 337 spontaneous shocks; 12 were inappropriate.
  • Supraventricular arrhythmias, particularly atrial fibrillation, caused 55% of inappropriate shocks.
  • Lead malfunction occurred in 4 patients; no mortality observed.

Conclusions:

  • Inappropriate shocks are frequent in patients with ICDs.
  • Atrial fibrillation is a primary cause of inappropriate shocks.
  • Shocks are most frequent in the initial six months post-implantation.