Related Experiment Videos
Sensitive and convenient high-performance liquid chromatographic method for the determination of mitomycin C in human
G Joseph1, W Biederbick, U Woschée
1Department of Pharmacology, University of Cologne, Germany.
Abstract:
An improved high-performance liquid chromatographic assay for the cytostatic drug mitomycin C in plasma is presented. The principal steps are precipitation of plasma proteins with acetonitrile, lyophilization of the supernatant and reversed-phase chromatography on a Hypersil ODS 5 microm column with 0.01 M NaH2PO4 buffer (pH 6.5)-methanol (70:30, v/v) in isocratic mode. At a flow-rate of 1.3 ml/min a column pressure of 180-220 bar resulted. Porfiromycin served as internal standard. UV detection was performed at 365 nm. Quantitation limit based on a coefficient of variation <10% in intra- and inter-day assay was 5 microg/l mitomycin C, detection limit based on a signal-to-noise ratio of 3 was 1 microg/l. Recovery was 100% and linearity was shown for the whole range of concentration (1-500 microg/l). None of the five drugs used during chemoembolisation interfered with the assay in vitro. The assay meets the requirements for pharmacokinetic studies of mitomycin C in patients as regards sensitivity and ease of use.