The Bain anaesthetic system. An assessment during controlled ventilation
Anaesthesia
|March 1, 1976
Summary
The Bain anaesthetic breathing system, a modified Mapleson D circuit, offers predictable carbon dioxide levels (PaCO2) with adjusted fresh gas flow in patients over 40 kg. This versatile system is suitable for all ages and various ventilation modes.
Area of Science:
- Anesthesiology
- Respiratory Physiology
- Medical Device Engineering
Background:
- The Mapleson D circuit is a common anesthetic breathing system.
- Controlled ventilation requires careful management of fresh gas flow.
- Optimizing anesthetic delivery and waste gas management is crucial.
Purpose of the Study:
- To evaluate the performance of the Bain anaesthetic breathing system (modified Mapleson D) under controlled ventilation.
- To determine the relationship between fresh gas inflow and PaCO2 levels.
- To assess the system's suitability as a universal breathing circuit.
Main Methods:
- Evaluation of the Bain circuit in over 140 patients undergoing anesthesia with controlled ventilation.
- Adjustment of fresh gas inflow based on patient body weight.
- Measurement of PaCO2 levels at different fresh gas flow rates.
Main Results:
- A predictable PaCO2 was achieved by adjusting fresh gas inflow relative to body weight in patients >40 kg.
- Mean PaCO2 was 40.8 mmHg at 70 ml/kg/min and 34.3 mmHg at 100 ml/kg/min fresh gas inflow.
- The circuit's length did not impact performance, even when doubled.
Conclusions:
- The Bain anaesthetic breathing system allows for predictable PaCO2 control with appropriate fresh gas flow adjustments.
- The system is versatile, lightweight, and can be used with simple ventilators for all patient ages.
- The Bain circuit demonstrates potential as a universal breathing system due to its performance and features.