An enhanced method for measuring cardiac output using Doppler color flow echocardiography
H Seo1, M Yamagishi, S A Haque
1Cardiology Division of Medicine, National Cardiovascular Center, Osaka, Japan.
Abstract:
An enhanced method for determining cardiac output using Doppler color flow imaging techniques to measure mitral orifice diameter was developed and validated in an experimental model and in clinical patients. In an in vitro circuit model, color jet width correlated well with actual orifice dimension from 12 to 24 mm (r = 0.99). In the clinical application, mitral valve area was calculated as a X b X pi/4 where a and b represent the width of the color flow stream in the mitral orifice just distal to the annulus in apical long-axis (short-diameter) and 4-chamber (90 degrees rotated, long-diameter) views, respectively. Cardiac output was then computed as the product of mitral valve area and time-velocity integral of transmitral flow from the same site. Cardiac output was also measured by thermodilution and conventional echocardiographic methods using diameters and time-velocity integrals from the left ventricular outflow tract. In 30 patients with nonvalvular heart disease, cardiac output measured by thermodilution ranged from 3.40 to 8.40 L/min. Cardiac output was determined in 28 of 30 patients (93%) by the Doppler color flow imaging technique; it ranged from 3.00 to 8.36 L/min and correlated well with thermodilution: y = 0.90x + 0.63, r = 0.91. Cardiac output was determined in 24 of 30 patients by the conventional left ventricular outflow method (80%). The cardiac output measured by the conventional method correlated less closely with thermodilution (r = 0.84), although there was no statistical difference in correlation coefficiencies between the 2 methods. These results indicate that the Doppler color flow imaging technique can be used to enhance the determination of cardiac output by echocardiography, particularly when the conventional method has resulted in technically inadequate recordings.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: