Potential risks and artifacts of magnetic resonance imaging of self-expandable esophageal stents
Related Experiment Video
Updated: Oct 31, 2026

Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
1Department of Gastroenterology, Netherlands Cancer Institute/Antoni van Leeuwenhoekhuis, Amsterdam.
Background:
Several types of coated and uncoated self-expandable stents composed of various metals are available for palliation of malignant esophageal stenoses and fistulas. We encountered poor visualization of the mediastinum and skeletal axis with magnetic resonance imaging in a patient with a Gianturco self-expandable stent.
Methods:
To evaluate potential problems, such as stent migration in the magnetic field of the magnetic resonance scanner (MRI) and artifacts in the images, we studied four types of expandable stents: the Ultraflex (titanium alloy), the covered Wall stent (nitinol), the Gianturco stent (Cook), and the modified Gianturco stent (Song)-the last two being made of stainless steel.
Results:
An appreciable attraction force and torque, as measured ex vivo by suspending the stent in a Perspex device, was found for both Gianturco stents. In particular, the Gianturco (Cook) stent is pulled toward the head with a force of 7g; however, it is uncertain whether this is a potential risk for dislodgement. In addition, gross artifacts in the images, studied by putting the stents in a cylindric phantom filled with a diluted gadolinium solution, made the magnetic resonance useless for imaging in a wide area around the Gianturco stents of stainless steel. No problems with magnetic resonance image quality were encountered for the titanium-based Ultraflex and Wall stents.
Conclusions:
Specific information on the type of stent is necessary before a magnetic resonance imaging examination is planned.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
09:40An Ivor Lewis Esophagectomy Designed to Minimize Anastomotic Complications and Optimize Conduit Function
Published on: April 17, 2020
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Esophageal Strictures-I: Introduction
Etiology
The primary cause of esophageal strictures is long-standing gastroesophageal reflux disease (GERD), accounting for about 70 to 80% of adult cases. Chronic acid reflux can lead to injury and scarring of the esophageal lining, culminating in...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations: