Idiosyncratic pacing with a commercial pacing system analyzer

S Zweibel1, S Goel, E Calderon

  • 1Pacemaker Center, Montefiore Medical Center, Bronx, New York 10467, USA.

Pacing and Clinical Electrophysiology : PACE
|February 10, 1998
PubMed
Summary

High output AAI pacing can cause unexpected ventricular capture with specific Medtronic pacing analyzer configurations. Physicians must recognize this anomaly to prevent misinterpretations during pacemaker testing.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Continuous bed motion versus step-and-shoot acquisition in LAFOV PET/CT: insights from multi-phantom and patient studies.

EJNMMI physics·2026
Same author

Uric acid and acute kidney injury in high-risk patients for developing acute kidney injury undergoing cardiac surgery: A prospective multicenter study.

Revista espanola de anestesiologia y reanimacion·2024
Same author

Adequacy and prognostic impact of treatment for severe exacerbation of chronic obstructive pulmonary disease.

Revista clinica espanola·2019
Same author

Importance of tissue sampling, laboratory methods, and patient characteristics for detection of Pneumocystis in autopsied lungs of non-immunosuppressed individuals.

European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology·2017
Same author

[Coronal dens in dente; clinical, radiographic and histopathological study].

Revista odontologica·2014
Same author

[Evolution of the intake and nutritional status of zinc, iron and copper in women undergoing bariatric surgery until the second year after surgery].

Nutricion hospitalaria·2013

Area of Science:

  • Cardiovascular Electrophysiology
  • Medical Device Technology

Background:

  • Medtronic AV Pacing System Analyzers (models 5311 and 5311B) are used for pacemaker testing.
  • AAI pacing is a mode where the atrium paces and senses, with the ventricle inhibited.

Purpose of the Study:

  • To report an idiosyncratic response observed with Medtronic AV Pacing System Analyzers during high output AAI pacing.
  • To alert implanting physicians to a potential artifact causing false ventricular capture.

Main Methods:

  • Utilized Medtronic AV Pacing System Analyzers (models 5311 unipolar and 5311B bipolar).
  • Performed high output AAI pacing with specific cable and adaptor configurations (unipolar analyzer with bipolar cables, bipolar analyzer with unipolar cables and 2734 adaptor).
  • Observed pacing response with displaced patient ground connection.

Related Experiment Videos

Main Results:

  • Ventricular capture was observed during high output AAI pacing when using the bipolar analyzer with unipolar cables and a 2734 adaptor, with the ground displaced.
  • Ventricular capture was also noted when using the unipolar analyzer with bipolar cables during high output AAI pacing.

Conclusions:

  • Specific configurations of Medtronic AV Pacing System Analyzers can lead to artifactual ventricular capture during high output AAI pacing.
  • Physicians must be aware of this phenomenon to avoid misinterpreting pacing results and drawing false conclusions about device function or patient condition.