Related Experiment Video
Updated: Aug 2, 2026

Measurement of Factor V Activity in Human Plasma Using a Microplate Coagulation Assay
Published on: September 9, 2012
A comparison between arterial- and venous-sampled activated clotting time measurements
E Zisman1, B Rozenberg, Y Katz
1Department of Anesthesiology, Rambam Medical Center, Haifa, Israel.
Abstract:
In order to compare activated clotting time (ACT) sampled from an arterial (heparin-flushed) line with the control, a venous (heparin-free) line, arterial and venous ACT values were assessed before and after cardiopulmonary bypass in 150 patients while undergoing open-heart surgery. ACT was measured by Hemochron 801 automatic analyzer. Baseline arterial ACT was significantly higher than baseline venous ACT (14%; p < 0.001, using one-way analysis of variance and Bonferroni multiple comparisons test). The differences between the values of arterial and venous ACT after protamine reversal, between arterial ACT at baseline and after protamine reversal, and between venous ACT at baseline and after protamine reversal were not statistically significant. We conclude that arterial-sampled ACT measurement is suitable and reliable for monitoring heparin reversal by protamine after cardiopulmonary bypass.
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis III: Interprofessional Care