Amplatz thrombectomy device: unexpected interaction with the Cragg endoprosthesis
H Schwarzenberg1, S Müller-Hülsbeck, M Heller
1Klinik für Radiologische Diagnostik, Christian-Albrechts-Universität zu Kiel, Arnold-Heller-Strasse 9, D-24105 Kiel, Germany.
Cardiovascular and Interventional Radiology
|March 28, 1998
Summary
Mechanical failure of the Amplatz thrombectomy device (ATD) occurred during graft occlusion treatment. Awareness of ATD interactions with graft struts is crucial for safe thrombectomy procedures.
Area of Science:
- Vascular Surgery
- Interventional Cardiology
Background:
- A 74-year-old woman underwent successful recanalization and implantation of a Dacron-covered Cragg endoprosthesis in the left superficial femoral artery for a 6-cm occlusion.
Observation:
- The endoprosthesis occluded 4.5 months post-discharge.
- An Amplatz thrombectomy device (ATD) was used for thrombectomy.
- The ATD experienced mechanical failure during the procedure, requiring complex extraction.
Findings:
- The ATD failure was likely caused by graft struts advancing into the device tip, breaking the drive shaft.
- Successful recanalization was ultimately achieved via percutaneous transluminal angioplasty and pharmacological thrombolysis.
Implications:
- Vascular surgeons and interventional cardiologists must be aware of potential interactions between thrombectomy devices and graft struts.
- This case highlights a rare complication that can occur during endovascular procedures involving thrombectomy devices and prosthetic grafts.