Design and biomechanical evaluation of a cementable endosteal blade implant
Abstract:
A cementable endosteal blade implant has been developed and evaluated. Inherent in the design are the following factors: minimization of stress concentrators, ease of implantation, and high resistance to loosening. Other potential advantages of this design as compared with conventional endosteal blade implants include reduced hazards of postoperative infection and reduced likelihood of metallic corrosion. Six conventional endosteal implants and six cementable implants were installed in steer mandibles using standard dental surgical techniques and Simplex-brand methyl methacrylate bone cement. They were loaded in uniaxial compresstion at a loading rate of 0.0122 in./min. the stiffness (S), deformation at 900-lb load (D900), proportional limil (PL), and load at 0.04 in. deformation (L0.04) were calculated for each test. The cementable design showed a twofold increase in stiffness, only 17% of the deformation at 900lb, more than twice the proportional limit, and at least twice the load at 0.04 in. deformation when compared with similar values for the conventional endosteal implants. This study reveals that, in addition to being unstable in bone, conventional endosteal implants are also unstable when used with acrylic bone cement. The new design should eliminate most of the problems associated with endosteal blade implantation.