Related Experiment Video
Updated: Aug 28, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Use of forward projection to correct patient motion during SPECT imaging
1Department of Medical Physics and Clinical Engineering, Royal Hallamshire Hospital, Sheffield, UK.
Abstract:
We have developed an iterative method to correct axial and tangential patient motion occurring during tomographic acquisition. The method uses axial images reconstructed from the uncorrected projection images, which are then forward projected to form a basis for registering the original planar images and, in the process, directly seeks to establish a consistent data set. Our method can be applied to all SPECT scans including myocardial and brain SPECT. We demonstrate that the method is capable of detecting and quantitatively correcting for complex motion in both axial and tangential directions. Results from phantom experiments show excellent resolution and contrast recovery after simulated movement in both the axial and tangential directions and initial results with clinical data sets are encouraging.
More Related Videos
11:29Real-time Video Projection in an MRI for Characterization of Neural Correlates Associated with Mirror Therapy for Phantom Limb Pain
Published on: April 20, 2019
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020