Deflection and stress distribution in three different IMZ abutment designs
C R Haganman1, D C Holmes, S A Aquilino
1Department of Restorative Dentistry, University of Colorado School of Prosthodontics, Denver 80262, USA.
Summary
The Abutment Complete (ABC) system showed significantly less stress and deflection in dental implants compared to the original threaded Intra-Mobile Element (IME) and Intra-Mobile Connector (IMC) systems. This suggests ABC may offer superior durability for prosthetic restorations.
Area of Science:
- Biomaterials Science
- Dental Implantology
- Mechanical Engineering
Background:
- Dental implant prosthetics require robust designs to withstand occlusal forces.
- Different prosthetic systems (IME, ABC, IMC) utilize varied designs for connecting abutments to implant fixtures.
- Understanding stress distribution is crucial for predicting the longevity and success of implant restorations.
Purpose of the Study:
- To compare stress distribution within the resin element and retaining screw for three IMZ prosthetic systems: IME, ABC, and IMC.
- To correlate stress distribution with prosthetic superstructure deflection under varying loads.
Main Methods:
- Finite element analysis (FEA) was used to create 3D models of implant-supported restorations for each system.
- Models simulated osseointegrated IMZ implant fixtures with identical crown and fixture representations.
- Non-axisymmetric loads (100 N and 500 N) were applied vertically and at a 45-degree angle to the occlusal surface.
Main Results:
- Deflections at the buccal cusp tip varied from 0.002 mm (ABC, 100 N vertical) to 0.802 mm (IME, 500 N, 45°).
- Maximum effective stresses in the retaining screw ranged from 129 MPa (ABC, 100 N vertical) to 1,315 MPa (IMC, 500 N, 45°).
- A direct correlation was found between peak screw stresses and superstructure deflection.
Conclusions:
- The Abutment Complete (ABC) system exhibited significantly lower deflections and stress concentrations compared to the IME and IMC systems.
- The Intra-Mobile Connector (IMC) and original threaded Intra-Mobile Element (IME) showed comparable, higher stress and deflection ranges.
- The ABC system may offer improved mechanical stability and longevity for dental prosthetic restorations.