Assay of S-ethyl-N-acetyl-l-cysteine in urine by high-performance liquid chromatography using post-column reaction
M Eskinja1, G Lamprecht, G Scherer
1Institut für Analytische Chemie der Universität Wien, Austria.
Abstract:
The assay of the ethyl chloride metabolite S-ethyl-N-acetyl-L-cysteine in human urine by HPLC is described. The compound is enriched by adsorption on a non-polar adsorbent of graphitized non-porous carbon, and then stripped from positively charged compounds by application onto a strong acid cation-exchanger. Subsequently, an enzymatic deacetylation is carried out and the acylase is removed by centrifugal ultrafiltration. Separation of the sample is performed by cation-exchange chromatography applying an eluent of a very low elution strength (diluted formic acid). In the column effluent S-ethyl-L-cysteine is derivatized by o-phthaldialdehyde and the reaction product is detected by fluorescence measurement. In human urine a detection limit in the low ppb range is achieved.