Rapid determination of p-aminohippuric acid in serum and urine by high-performance liquid chromatography
Related Experiment Video
Updated: Oct 23, 2026

Biochemical Measurement of Neonatal Hypoxia
Published on: August 24, 2011
R Marsilio1, R Dall'Amico, G Montini
1Department of Pediatrics, University of Padua, Italy.
Abstract:
We report a simple and rapid HPLC procedure for measuring p-aminohippuric acid in serum and urine. After deproteinization with acetonitrile and addition of p-aminobenzoic acid as an internal standard, the chromatographic run is performed on a C18 column with the absorbance detector set at 275 nm. The separation is carried out in 10 min at a flow-rate of 1.0 ml/min with a mobile phase composed of 7 mmol/l 1-decanesulfonic acid, pH 3.70, and acetonitrile (82:18, v/v). The relationship between p-aminohippuric acid concentration and the p-aminohippuric acid/internal standard peak area is linear up to 100 microg/ml. Within-run precision measured at three different p-aminohippuric acid concentrations ranges from 1.73 to 1.98% in serum and from 0.72 to 1.32% in urine. Between-run precision varies from 1.80 to 4.06% in serum and from 1.05 to 2.66% in urine. Analytical recovery is between 98.26 and 99.44% in serum and from 99.57 to 100.45% in urine. The method is very simple, sensitive, and requires small volumes of sample for the assay (100 microl). Therefore, it could be a useful tool for the analysis of p-aminohippuric acid in the evaluation of renal plasma flow.
More Related Videos
11:29HPLC-based Assay to Monitor Extracellular Nucleotide/Nucleoside Metabolism in Human Chronic Lymphocytic Leukemia Cells
Published on: July 20, 2016
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018