Discrepancy between stent deployment and balloon size used assessed by intravascular ultrasound
M R van Sambeek1, A Qureshi, W van Lankeren
1University Hospital, Rotterdam-Dijkzigt, The Netherlands.
Objectives:
This study was designed to assess the discrepancy in stent deployment seen on intravascular ultrasound and its relation to the balloon size selected for stent delivery.
Design:
Prospective study.
Materials And Methods:
The study group comprised 27 patients treated using a stent (n = 18) or stent-graft combination (n = 9). Following angiographically optimal stent deployment (< 10% residual stenosis) intravascular ultrasound was used to compare the smallest intra-stent lumen area with measurements at both stent edges and the lumen area of the proximal and distal reference sites.
Results:
In 14 of the 27 stents the intra-stent dimension was the same as the dimension of the stent edge (difference < or = +/- 10%). Of the remaining stents the intra-stent dimension was smaller (difference > 10%) than the proximal stent edge in seven stents (range 11-39%), smaller than the distal stent edge in three stents (range 11-20%) and smaller than both stent edges in three stents (range 12-37%). Both in patients treated with a stent or stent-graft combination, the resulting smallest intra-stent lumen area was smaller than the balloon size used (mean difference 32% and 42%, respectively) and smaller than the mean lumen area of the reference sites (mean difference 25% and 23%, respectively).
Conclusion:
This intravascular ultrasound study shows a discrepancy between intra-stent lumen area, the area of the stent edges and the balloon size used.