Related Experiment Videos
Application of nitrite reductase from Alcaligenes faecalis S-6 for nitrite measurement
1Research Center for Advanced Science and Technology, University of Tokyo, Japan.
Abstract:
The enzymatic reaction of nitrite reductase (NIR) from Alcaligenes faecalis S-6 was applied to the measurement of nitrite. NIR was immobilized on the surface of a gold electrode using filter paper and a dialysis membrane, and used as a working electrode in a three-electrode system. Amperometric methods were applied using NIR and the electron mediator 1-methoxy PMS (1-methoxy-5-methylphenazinium methylsulfate). The decrease in cathodic current showed a correlation to nitrite concentration over the range 0-1 mg/l. Measurements using a batch-flow type system gave a lower detection limit of 0.01 mg/l. This is sufficient for the detection of nitrite in natural waters.