Related Experiment Video
Updated: Aug 4, 2026

MRI and PET in Mouse Models of Myocardial Infarction
Published on: December 19, 2013
[18F]fluorodeoxyglucose single photon emission computed tomography: can it replace PET and thallium SPECT for the
G Srinivasan1, A N Kitsiou, S L Bacharach
1Cardiology Branch, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD 20892-1650, USA.
Background:
New high-energy collimators for single photon emission computed tomography (SPECT) cameras have made imaging of positron-emitting tracers, such as [18F]fluorodeoxyglucose (18FDG), possible. We examined differences between SPECT and PET technologies and between 18FDG and thallium tracers to determine whether 18FDG SPECT could be adopted for assessment of myocardial viability.
Methods And Results:
Twenty-eight patients with chronic coronary artery disease (mean left ventricular ejection fraction [LVEF]=33+/-15% at rest) underwent 18FDG SPECT, 18FDG PET, and thallium SPECT studies. Receiver operating characteristic curves showed overall good concordance between SPECT and PET technologies and thallium and 18FDG tracers for assessing viability regardless of the level of 18FDG PET cutoff used (40% to 60%). However, in the subgroup of patients with LVEF< or =25%, at 60% 18FDG PET threshold value, thallium tended to underestimate myocardial viability. In a subgroup of regions with severe asynergy, there were considerably more thallium/18FDG discordances in the inferior wall than elsewhere (73% versus 27%, P<.001), supporting attenuation of thallium as a potential explanation for the discordant observations. When uptake of 18FDG by SPECT and PET was compared in 137 segments exhibiting severely irreversible thallium defects (scarred by thallium), 59 (43%) were viable by 18FDG PET, of which 52 (88%) were also viable by 18FDG SPECT. However, of the 78 segments confirmed to be nonviable by 18FDG PET, 57 (73%) were nonviable by 18FDG SPECT (P<.001).
Conclusions:
Although 18FDG SPECT significantly increases the sensitivity for detection of viable myocardium in tissue declared nonviable by thallium (to 88% of the sensitivity achievable by PET), it will occasionally (27% of the time) result in falsely identifying as viable tissue that has been identified as nonviable by both PET and thallium.
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT