Low-vacuum, low-voltage scanning electron microscopy of poly(methyl methacrylate) intraocular lenses
Purpose:
To evaluate the usefulness of low-vacuum, low-voltage scanning electron microscopy (LVSEM) in examining the surfaces of intraocular lenses (IOLs) without preparation.
Setting:
In vitro laboratory experiment.
Methods:
Six commercially available IOLs made of poly(methyl methacrylate) (PMMA) and one explanted PMMA IOL were evaluated with LVSEM without sputter coating. Sham surgical maneuvers were performed on some IOLs and the surface changes observed.
Results:
Smooth surfaces of IOL optics and haptics, relatively sharp edges of optic holes, and irregularities at the optic/haptic junction were clearly observed. Crystalline deposits were seen on the surfaces after the IOLs were dipped and dried in saline solution. Scratches and marks were observed after they were grasped with a forceps.
Conclusions:
Low-vacuum, low-voltage scanning electron microscopy is a time-saving procedure that can be used by inexperienced investigators. The resolution under relatively low magnification appeared to be satisfactory for evaluating the IOL surface.