Related Experiment Video
Updated: Aug 9, 2026

Whole Genome Sequencing of Candida glabrata for Detection of Markers of Antifungal Drug Resistance
Published on: December 28, 2017
Evaluation of commercial systems for the identification of clinical yeast isolates
Abstract:
The Analytab Products Inc. (API), Micro-Drop (MD), and Uni-Yeast-Tek (UYT) systems for the presumptive identification of common clinical yeast isolates were compared with the oxidation-fermentation (OF) and a conventional procedure. With 229 coded isolates, the identification accuracies were API 94, MD 83, OF 82, and UYT 99%. The API system required the greatest technical ability. The MD materials were prone to malfunction. OF media, if incubated beyond 14 days, gave an accuracy of 87%, but this offered no advantage over the conventional procedure. The UYT system was the easiest to use.
Related Concept Videos
Methods of Classification and Identification
Rapid Identification of Pathogens
Automated Microbial Diagnostics