Background:
None of the presently used airway devices are ideal regarding ease of insertion, alignment with the laryngeal inlet, and provision of a high-pressure seal from the environment. The purpose of this study was to determine, in awake volunteers, the performance of a new ventilatory device, the glottic aperture seal airway, regarding ease of insertion, alignment with the laryngeal inlet, and forced exhalation seal pressure (PFES).
Methods:
The glottic aperture seal airway consists of a curved tubular component that ends in the middle of an elliptical foam cushion glottic component. The posterior surface of the foam has a curved flexible plastic backing, which imparts a 60 degree angle between the proximal half and the distal half of the foam cushion. When the glottic aperture seal airway is properly in situ in a supine patient, the proximal half of the foam cushion is opposite the laryngeal inlet. The posterior surface of the plastic backing has a balloon attached to it. Inflation of the balloon presses the ventilation hole and foam cushion up against the laryngeal inlet, thereby creating a seal from the environment. Using the laryngeal mask airway as a control device, the glottic aperture seal airway was tested for time and ease of insertion, fiberoptic alignment with the laryngeal inlet, and PFES in 18 lightly sedated and locally anesthetized volunteers.
Results:
The glottic aperture seal and laryngeal mask airways were inserted with equal ease and speed. The fiberoptic alignment with the larynx was excellent for both the glottic aperture seal and laryngeal mask airways. In all volunteers, the mean +/- SD PFES values at 0-, 10-, 20-, 30-, and 40-ml balloon inflation volumes of the glottic aperture seal airway were 23.4 +/- 11.8, 29.6 +/- 12.4, 42.7 +/- 12.5, 56.9 +/- 5.6, and 60 +/- 0 cm H2O, respectively; the PFES at > or = 20 ml balloon inflation volume of the glottic aperture seal airway was significantly greater than with the laryngeal mask airway (19.4 +/- 6.7 cm H2O, P < 0.01). A PFES of > or =60 cm H2O was achieved with the glottic aperture seal airway in all volunteers (n = 2 at 10 ml, n = 3 at 20 ml, n = 9 at 30 ml, and n = 4 at 40 ml). The glottic aperture seal airway did not cause any trauma.
Conclusion:
In awake volunteers, the glottic aperture seal and laryngeal mask airways were equally easy to insert and position. The glottic aperture seal airway was capable of achieving a higher PFES than the laryngeal mask airway.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques