Cranial bone flap fixation clamps: compatibility at MR imaging

F G Shellock1, V J Shellock

  • 1Department of Radiology, School of Medicine, University of Southern California, Los Angeles, USA.

Radiology
|June 4, 1998
PubMed
Summary

Metallic cranial bone flap fixation clamps are safe for 1.5-T MRI scans. Testing showed no significant magnetic attraction, minimal heating, and minor artifacts, indicating patient safety during magnetic resonance imaging.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Assessment of MRI safety issues for stainless steel sutures used for microtia reconstruction.

Journal of plastic, reconstructive & aesthetic surgery : JPRAS·2018
Same author

Physiological Benefits of Warm-Up.

The Physician and sportsmedicine·2016
Same author

Kinematic Magnetic Resonance Imaging for Evaluation of Patellar Tracking.

The Physician and sportsmedicine·2016
Same author

In vitro evaluation of MR imaging issues at 3T for aneurysm clips made from MP35N: Findings and information applied to 155 additional aneurysm clips.

AJNR. American journal of neuroradiology·2009
Same author

Kinematic magnetic resonance imaging of the effect of bracing on patellar position: qualitative assessment using an extremity magnetic resonance system.

Journal of athletic training·2006
Same author

Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.

AJNR. American journal of neuroradiology·2006

Area of Science:

  • Neurosurgery
  • Radiology
  • Biomedical Engineering

Background:

  • Cranial bone flap fixation is crucial for neurosurgical procedures.
  • Assessing the safety of metallic implants in magnetic resonance imaging (MRI) is essential.

Purpose of the Study:

  • To evaluate the magnetic compatibility of metallic cranial bone flap fixation clamps.
  • To determine potential risks associated with 1.5-Tesla (T) MRI in patients with these implants.

Main Methods:

  • Metallic cranial bone flap fixation clamps were subjected to standardized 1.5-T MRI.
  • Tests included assessment of magnetic field attraction, heating effects, and image artifact generation.

Main Results:

  • No significant magnetic field attraction was observed.

Related Experiment Videos

  • Minor heating effects were recorded.
  • Image artifacts resulted in a relatively small signal void.
  • Conclusions:

    • Metallic cranial bone flap fixation clamps demonstrate compatibility with 1.5-T MRI systems.
    • These implants do not pose a significant risk to patients undergoing 1.5-T MRI procedures.