Detection of persistent vegetative bacteria and amplified viral nucleic acid from in-use testing of gastrointestinal

A K Deva1, K Vickery, J Zou

  • 1Department of Infectious Diseases, University of Sydney, Australia.

Insights

Thorough manual cleaning of gastrointestinal endoscopes is crucial. Inadequate cleaning can lead to persistent bacteria and viruses like hepatitis B virus (HBV) and hepatitis C virus (HCV), increasing infection risk.

Area of Science:

  • Infection Control
  • Gastroenterology
  • Microbiology

Background:

  • Hospital-acquired infections linked to endoscope decontamination failures.
  • Gastrointestinal endoscopes require rigorous cleaning protocols to prevent pathogen transmission.

Purpose of the Study:

  • To evaluate the effectiveness of current gastrointestinal endoscope decontamination procedures.
  • To identify potential breaches in cleaning protocols that could lead to patient infections.

Main Methods:

  • Collected 123 endoscope channel specimens before, during, and after decontamination.
  • Used broth and plate cultures to detect vegetative bacteria.
  • Employed polymerase chain reaction (PCR) to detect hepatitis B virus (HBV), hepatitis C virus (HCV), and human immunodeficiency virus (HIV-1) nucleic acids.

Main Results:

  • 34% of samples showed bacterial growth after glutaraldehyde disinfection due to manual cleaning errors; corrected after intervention.
  • Peracetic acid failed to sterilize in one unit, with 15% of samples showing bacterial growth.
  • HCV and HBV DNA detected in endoscopes used on infected patients during improper cleaning; viral nucleic acid was not detected after proper cleaning.

Conclusions:

  • Effective manual cleaning is essential before chemical disinfection of endoscopes.
  • Bacterial detection in endoscope channels signals decontamination failures.
  • Inadequate cleaning may leave persistent viral nucleic acid (HCV, HBV) on endoscopes.