Radiation shielding implications of computed tomography scatter exposure to the floor
1Department of Diagnostic Radiology, Mayo Clinic and Foundation, Rochester, MN 55905, USA.
Abstract:
This article investigates whether CT scanners used primarily for either head or body work have scattered radiation levels that may impact shielding considerations. Several 35 x 43 cm film cassettes were used to cover the floor area around a General Electric Hi-Speed Advantage CT unit. Regions of maximum exposure were identified (visually with sensitometric confirmation). An ion chamber was used to obtain radiation exposure levels for typical head and body protocols. Greater scattered radiation was found behind the gantry for head scanning (20 vs. 16 microGy for a body scan) while greater scatter was observed in front of the scanner for body scans (46 vs. 31 microGy for the head scan). The resulting annual doses to the floor from typical workloads (assuming 32 cases a day) can be as great as 0.63 Gy for head protocols or 1 Gy from body work. Care may be needed to assure adequate shielding for floor areas near CT scanners specializing in either head or body work. Also, while increased throughput with helical scanners is currently x-ray tube heat limited, shielding plans should allow for enhanced heat capacity (and greater throughput) in future generation scanners.
Related Concept Videos
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Biological Effects of Radiation
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography