Sensitivity to metals in 40 patients with failed hip endoprosthesis
V Milavec-Puretić1, D Orlić, A Marusić
1Department of Dermatology, Zagreb University Hospital Center and School of Medicine, Croatia.
Archives of Orthopaedic and Trauma Surgery
|August 26, 1998
Summary
Metal hypersensitivity was investigated in revision hip arthroplasty patients. Patch tests showed metal sensitivity in some, but it likely doesn't significantly impact implant loosening or revision safety.
Area of Science:
- Orthopedic Surgery
- Immunology
- Biomaterials Science
Background:
- Implant loosening is a common complication in hip arthroplasty.
- Concerns exist regarding metal hypersensitivity and its role in implant failure.
- Revision hip arthroplasty involves complex patient cases and material considerations.
Purpose of the Study:
- To investigate the prevalence of metal hypersensitivity in patients undergoing revision hip arthroplasty.
- To determine if metal sensitivity correlates with tissue reaction or implant loosening.
- To assess the safety of stainless steel endoprostheses in repeated revision surgeries.
Main Methods:
- Studied 40 patients undergoing revision hip arthroplasty.
- Conducted standard cutaneous patch testing for metal sensitivity (chromium, nickel, cobalt, rust, scrapings).
- Compared clinical and histological data between patients with positive and negative patch tests.
Main Results:
- Nine out of 40 patients (22.5%) exhibited positive patch test reactions to various metal allergens.
- No significant differences were observed between positive and negative patch test groups regarding age, sex, diagnosis, revision history, implant function, or tissue histology.
- Circulating immunocomplex levels also did not differ between the groups.
Conclusions:
- Stainless steel endoprostheses appear safe for use in repeated revision hip arthroplasty.
- Metal hypersensitivity is likely not a significant etiological factor in the loosening of hip endoprostheses.
- Further research may explore other mechanisms contributing to implant failure in revision cases.