Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve
D M Philbin1, M A Marieb, K H Aithal
1Section of Cardiovascular Medicine, Yale University School of Medicine, New Haven, Connecticut 06511, USA.
Pacing and Clinical Electrophysiology : PACE
|October 30, 1998
Abstract:
Although the potential for transcutaneous electronic nerve stimulation (TENS) units to interfere with appropriate function of cardiac pacemakers is well documented, an adverse interaction between a TENS unit and an ICD has never been reported. We describe a patient in whom a TENS unit created an electrical artifact that was interpreted by a transvenous ICD as ventricular fibrillation, leading to the delivery of inappropriate therapy. TENS units should be used with caution in patients with ICDs.