Quantitative determination of Ecteinascidin 743 in human plasma by miniaturized high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry
Related Experiment Video
Updated: Jul 28, 2026

Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
H Rosing1, M J Hillebrand, J M Jimeno
1Department of Pharmacy and Pharmacology, Slotervaart Hospital/The Netherlands Cancer Institute, Amsterdam, The Netherlands.
Abstract:
A method was developed for the bio-analysis of Ecteinascidin 743 (ET-743) using miniaturized liquid chromatography (LC) coupled to an electrospray ionization sample inlet (TurbolonSpray) and two quadrupole mass analyzers (LC/ESI-MS/MS). Solid-phase extraction was used as a sample pretreatment procedure. Ecteinascidin 743 is a very potent anticancer compound and is administered in microgram m-2 dosages, which demands special requirements in terms of sensitivity for the analytical method supporting clinical pharmacokinetic studies. Using conventional LC/UV, a lower limit of quantitation (LLQ) of 1 ng ml-1 plasma was reached using a 500 microliters sample volume, but LC/ESI-MS/MS permitted an LLQ of 10 pg ml-1. The latter method was found to be accurate and precise, and provided a broad linear concentration range of 0.010-2.50 ng ml-1.