Related Experiment Videos
Patient radiation doses during cardiac catheterization procedures
S Betsou1, E P Efstathopoulos, D Katritsis
1Medical Physics Department, School of Medicine, University of Patras, Greece.
The British Journal of Radiology
|December 16, 1998
Summary
This study measured patient radiation doses during cardiac procedures like coronary angiography and PTCA using TLDs and DAP meters. Effective doses ranged from 5.6 to 13.0 mSv, with cinefluorography contributing significantly to patient exposure.
Area of Science:
- Medical Physics
- Radiology
- Cardiology
Background:
- Patient radiation dose assessment is crucial in interventional cardiology procedures.
- Cardiac procedures like coronary angiography and percutaneous transluminal coronary angioplasty (PTCA) involve significant radiation exposure.
- Accurate dose monitoring is essential for patient safety and optimizing radiation protection strategies.
Purpose of the Study:
- To determine the radiation dose received by patients undergoing cardiac procedures.
- To quantify the contribution of different fluoroscopic techniques to the overall patient dose.
- To establish a conversion factor for estimating effective dose from dose-area product (DAP) measurements.
Main Methods:
- Utilized calibrated Thermoluminescent Dosemeters (TLDs) to measure skin dose at four anatomical locations.
- Employed a Dose-Area Product (DAP) meter to assess cumulative radiation exposure.
- Estimated the DAP to effective dose conversion factor using a Rando phantom.
Main Results:
- Cinefluorography contributed more to the total DAP than fluoroscopy.
- A DAP to effective dose conversion factor of 0.183 mSv Gy⁻¹ cm⁻² was determined.
- Mean effective doses varied by procedure: 5.6 mSv (coronary angiography), 6.9 mSv (PTCA), 9.3 mSv (coronary angiography + ad hoc PTCA), 9.0 mSv (PTCA + stent), and 13.0 mSv (coronary angiography + ad hoc PTCA + stent).
Conclusions:
- The study successfully assessed patient effective doses during various cardiac interventions.
- The findings provide valuable data for radiation risk management in interventional cardiology.
- Accurate dose estimation is feasible using TLDs and DAP measurements with established conversion factors.