Extra-cellular volume estimation by electrical impedance--phase measurement or curve fitting: a comparative study

H G Goovaerts1, T J Faes, G W de Valk-de Roo

  • 1Department of Clinical Physics and Informatics, University Hospital Vrije Universiteit, Amsterdam, The Netherlands. hg.goovaerts@azvu.nl

Physiological Measurement
|December 24, 1998
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Differences in pharmacological activities of the antioxidant flavonoid monoHER in humans and mice are caused by variations in its metabolic profile.

Clinical pharmacology and therapeutics·2011
Same author

Anomalous diffraction by arbitrarily oriented ellipsoids: applications in ektacytometry.

Applied optics·2010
Same author

Light scattering by red blood cells in ektacytometry: Fraunhofer versus anomalous diffraction.

Applied optics·2010
Same author

Assessing effects of PEEP and global expiratory lung volume on regional electrical impedance tomography.

Journal of medical engineering & technology·2009
Same author

Interactions between different EEG frequency bands and their effect on alpha-fMRI correlations.

NeuroImage·2009
Same author

Characteristics of the parallel-plate flow chamber for mechanical stimulation of bone cells under microgravity.

Journal of gravitational physiology : a journal of the International Society for Gravitational Physiology·2009

This study introduces a novel Bode analysis method for accurately measuring body fluid shifts. The new phase measurement technique offers a more precise estimation of extracellular volume (ECV) compared to traditional Cole-Cole extrapolation.

Area of Science:

  • Biomedical Engineering
  • Physiology
  • Medical Physics

Background:

  • Assessing body fluid distribution between intracellular and extracellular spaces is crucial for clinical diagnostics.
  • Traditional multifrequency impedance analysis relies on Cole-Cole extrapolation, which can be prone to inaccuracies.
  • Accurate estimation of extracellular volume (ECV) is essential for fluid management and patient monitoring.

Purpose of the Study:

  • To present a new approach for determining intra- and extra-cellular conduction using Bode analysis.
  • To enable continuous monitoring of the relationship between intra- and extra-cellular conduction via phase measurement.
  • To improve the accuracy of extracellular volume (ECV) estimation compared to existing methods.

Main Methods:

  • Multifrequency impedance measurements were conducted across a frequency range of 5 kHz to 1 MHz.

Related Experiment Videos

  • A novel method based on Bode analysis was developed to estimate intra- and extra-cellular conduction.
  • Extracellular volume (ECV) was estimated using a two-frequency measurement (4 kHz and 64 kHz) and compared with Cole-Cole extrapolation.
  • Main Results:

    • The proposed Bode analysis method allowed for ECV estimation with significantly higher accuracy than Cole-Cole extrapolation.
    • The new method demonstrated a strong correlation (0.90) with the reference sodium bromide method for ECV determination.
    • The average error in ECV estimation using the new method was -3.6%, compared to 13.2% for Cole-Cole extrapolation.

    Conclusions:

    • Bode analysis offers a more accurate and reliable method for estimating extracellular volume (ECV).
    • Continuous monitoring of body fluid shifts is achievable through simple phase measurements without complex curve fitting.
    • This technique provides a valuable tool for clinical assessment of body fluid status.