Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings
D V Exner1, J M Rothschild, S Heal
1Division of Cardiology, Foothills Hospital, Calgary, Alberta, Canada.
Summary
Contemporary unipolar leads show similar sensing to bipolar leads, but atrial unipolar leads are more prone to myopotential inhibition. Optimal programming reduces oversensing risks with unipolar leads.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Bipolar lead use is rising due to oversensing issues with older unipolar systems.
- Limited data exists on modern unipolar leads with improved design and flexibility.
Purpose of the Study:
- To compare sensing characteristics of contemporary unipolar and bipolar leads.
- To evaluate oversensing and myopotential inhibition in different lead polarities.
Main Methods:
- Randomized crossover study comparing unipolar and bipolar sensing in 34 patients.
- Intracardiac electrogram amplitudes, pacing/sensing thresholds, and provocative maneuvers were assessed.
- Optimal sensitivity settings were determined and validated with ambulatory ECG monitoring.
Main Results:
- Atrial unipolar leads showed significantly higher myopotential inhibition (50% vs. 5%) at higher sensitivities.
- Oversensing occurred in 6% of atrial and 8% of ventricular unipolar leads at optimal settings, versus none in bipolar leads.
- Undersensing was more frequent in atrial unipolar leads (29% vs. 6%).
Conclusions:
- Atrial unipolar leads can be prone to myopotential inhibition and oversensing, though optimal programming mitigates risks.
- Unipolar ventricular leads may be a viable option given concerns about bipolar lead longevity.
Related Concept Videos
Pulse rhythm
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Potentiometry: Overview
Potentiometry is an analytical technique that measures the potential difference between two electrodes in an electrochemical cell without drawing any significant current that could alter the solution's composition. This method employs an indicator electrode, which exchanges electrons with the analyte solution, and a reference electrode with a constant potential. Each electrode is immersed in a solution comprised of two half-cells. In a conventional setup, the reference electrode serves as the...
Potentiometry: Membrane Electrodes
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at the...