Are serum cortisol measurements by immunoassays reliable?: A case series

Nowreen Haq1, Katherine A Araque2, Anastasia L Gant Kanegusuku3

  • 1Division of Endocrinology and Metabolism. National Institute of Diabetes, Digestive and Kidney Diseases (NIDDK). National Institutes of Health, Clinical Center. Bethesda, MD, United States of America.

Medical Research Archives
|July 22, 2021
PubMed

Insights

Automated immunoassays may provide inaccurate thyroid hormone and serum cortisol measurements. High or low levels of binding globulins, such as thyroxine-binding globulin and corticosteroid binding globulin, can affect assay reliability.

Area of Science:

  • Clinical Chemistry
  • Endocrinology
  • Assay Development

Background:

  • Automated immunoassays (IA) are widely used for measuring various analytes, including hormones.
  • Previous studies have indicated unreliable thyroid hormone measurements with IA in cases of abnormal thyroxine-binding globulin (TBG) levels.
  • Many hormones circulate bound to specific binding proteins, which can potentially interfere with immunoassay accuracy.

Purpose of the Study:

  • To investigate the accuracy of automated immunoassays for serum cortisol measurement.
  • To compare IA-derived cortisol concentrations with those obtained by a reference method (liquid chromatography-tandem mass spectrometry).
  • To identify potential factors affecting IA accuracy for serum cortisol.

Main Methods:

  • Comparison of serum cortisol measurements obtained by routine automated immunoassays (IA) and liquid chromatography-tandem mass spectrometry (LC-MS/MS).
  • Analysis of preliminary data from a series of cases.
  • Evaluation of the impact of corticosteroid binding globulin (CBG) levels on IA results.

Main Results:

  • Automated immunoassays consistently reported falsely low serum cortisol concentrations, with an average underestimation of 27%.
  • Preliminary findings suggest that high concentrations of corticosteroid binding globulin (CBG) interfere with IA measurements of serum cortisol.
  • This interference leads to inaccurate low cortisol readings.

Conclusions:

  • Routine automated immunoassays may not provide accurate serum cortisol measurements, particularly in the presence of elevated corticosteroid binding globulin.
  • The findings highlight a potential limitation of IA for cortisol analysis, similar to issues observed with thyroid hormone assays.
  • Further validation using LC-MS/MS is recommended for accurate cortisol quantification in clinical settings, especially when CBG levels are abnormal.