Nationwide quality control trial on lymphocyte immunophenotyping and flow cytometer performance in Italy
1Renal Transplant Unit, Niguarda-Cà Granda Hospital, Milan, Italy.
Insights
This nationwide quality control trial assessed cellular immunophenotyping and flow cytometer performance. Instrument variability was high, highlighting the need for improved operator training and technical guidelines.
Area of Science:
- Immunology
- Biotechnology
- Medical Diagnostics
Background:
- The Italian Society for Cytometry (GIC) initiated a nationwide quality control trial in 1991.
- The study focused on cellular immunophenotyping and flow cytometer performance across Italy.
Purpose of the Study:
- To evaluate flow cytometer instrument performance using calibrated fluorescence microbeads.
- To correlate instrument performance with the measurement of lyophilized lymphocyte surface immunofluorescence staining.
- To assess the skill of cytometer operators in a large-scale setting.
Main Methods:
- A nationwide quality control trial involving 306 laboratories and 350 instruments.
- Use of calibrated fluorescence microbeads for assessing FITC threshold, linearity, CVs, and resolution.
- Analysis of lyophilized human lymphocytes stained with various antibodies (FITC and FITC/PE).
Main Results:
- Significant heterogeneity in instrument performance, particularly in FITC thresholds and coefficients of variation.
- Low FITC thresholds correlated with good resolution for low-intensity stainings.
- Outlier frequencies in lymphocyte analysis ranged from 3.2% to 19.7%, with variations based on antibody and marker combinations.
Conclusions:
- The study design was effective for evaluating instrument performance and operator skills.
- Identified practical challenges among participants indicate a need for enhanced educational programs and technical guidelines.
- Instrument performance variability underscores the importance of standardized quality control in flow cytometry.
Abstract:
The Italian Society for Cytometry (Gruppo Italiano di Citometria, GIC) promoted this nationwide large-scale quality control trial on cellular immunophenotyping and flow cytometer performance in 1991. The aim of this independent study was to evaluate instrument performance with calibrated fluorescence microbeads (minimum threshold for FITC, linearity, coefficients of variation and resolution indexes for fluorescence), and to correlate it to the measurement of lyophilized lymphocyte surface immunofluorescence staining with a wide spectrum of antibodies (percentage of positive cells and fluorescence mean intensity). A single send-out was made to 306 laboratories with 350 instruments throughout Italy. Each kit included anonymous vials containing premixed calibrated microbeads, lyophilized human lymphocytes and small aliquots of conjugated monoclonal antibodies, for both single FITC and double FITC/PE staining. Participants were also asked to use their own anti-CD4 monoclonal. A valid answer was returned by 209 laboratories with 221 instruments. Gating was not an assay variable. The minimum sensitivity threshold for FITC ranged from 26.6 to 11,293 MESF, with marked instrument heterogeneity as far as the relationship between FITC threshold and coefficients of variation for fluorescence was concerned. A low FITC threshold also correlated with a good resolution index for low intensity stainings. No performance comparisons among instrument brands and models were made. The lyophilized lymphocyte analysis showed an overall frequency of outliers ranging from 3.2% to 19.7%, with a maximum for CD2 FITC (19.7%) and CD3 + HLA-DR + (12.1%). A strongly negative relationship was evident between the FITC threshold and the number of CD2+ cells, whereas the sensitivity threshold had virtually no effects on the measured level of higher antigen density markers like CD4 and many others. Absolute fluorescence intensity measurements of CD4+ and CD19+ were also made. This study design proved valid and suitable for a large-scale evaluation of both instrument performance and cytometer operators' skill. A number of practical problems were identified among participants, thus stressing the need for more effective educational programmes and technical guidelines.