输液的流量率偏差:输液组的缺陷使发生故障,并导致相当大的精度偏差
Axel Hult1,2,3,4,5, Irina Zholobova1,2,3,4,5, Emelie Bäcklin1,2,3,4,5
1Karolinska University Hospital Huddinge, Stockholm, Sweden (Hult, Zholobova, Bäcklin); Astrid Lindgren Children's Hospital, Karolinska University Hospital, Stockholm, Sweden (Nydert).
概括
缺陷的输液套件可能会导致体积输液输送的药物少于规定的,没有任何警报. 这种低输液错误的风险威胁到患者的安全和药物准确的输送.
科学领域:
- 医疗器械 医疗器械
- 生物医学工程 生物医学工程
- 患者安全 患者安全
背景情况:
- 卷度输液对于药物输送至关重要.
- 流速错误 (过量/不足输注) 是已知的问题.
- 最近在瑞典一家医院报告了输液不足的错误.
研究的目的:
- 调查体积输液中输入不足的错误.
- 确定有缺陷的输液装置对准确性的影响.
- 评估未被检测到的药物不足的风险.
主要方法:
- 使用重力测量方法进行流速精度测试的实验研究.
- 操纵和临床使用的有缺陷的输液组之间的比较.
- 专注于具有可见片段延长的输液套.
主要成果:
- 缺陷的输液套件导致了显著的精度偏差.
- 不准确地报告了比实际交付更高的注入量.
- 由于降低了药物输出,没有引发任何警报.
- 长长的套件导致大量的输液不足,不反映在显示器上.
结论:
- 缺陷的输液套件对患者的安全构成严重风险.
- 输液不足的错误可能不会被和临床医生发现.
- 确保输液组的完整性对于准确的药物管理至关重要.
相关概念视频
Pipe Flowrate Measurement
699
In pipe flow measurement, orifice, nozzle, and Venturi meters are commonly used to determine fluid flowrates by constricting the flow area, which increases fluid velocity and reduces pressure. This pressure difference, governed by Bernoulli's principle and adjusted for real-world conditions, is essential for calculating flowrate. Each meter type is suited to specific applications based on accuracy, efficiency, and compatibility with various flow conditions.
The orifice meter is a simple,...
The orifice meter is a simple,...
699
Errors occurring during blood pressure monitoring
730
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
730
Pipe Flowrate Measurement: Problem Solving
545
A spray tank system is engineered to uniformly distribute a pest-control liquid across plants by using a pressurized mechanism. The tank, pressurized to 150 kPa, holds the pesticide at a height of 0.80 meters. Liquid flows from the tank through a 1.9 meter pipe with a diameter of 0.015 meters, angled at 0.698 radians, ultimately reaching a 0.007 meter nozzle that sprays the pesticide. Accurate calculation of the system's flow rate is crucial to ensure uniform application, and this is...
545
Uncertainty in Measurement: Accuracy and Precision
73.7K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.
73.7K
Single Pipe Systems
143
In pipe flow analysis, problems are typically categorized into three types — Type I, Type II, and Type III — based on the known parameters and the desired outcome. Each type of problem addresses specific engineering requirements using fluid properties, pipe characteristics, and operational conditions.
In a Type I problem, fluid properties (density and viscosity), pipe characteristics (including diameter, length, and surface roughness), and the flow rate or average velocity are...
In a Type I problem, fluid properties (density and viscosity), pipe characteristics (including diameter, length, and surface roughness), and the flow rate or average velocity are...
143
Glassware Calibration
240
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
240